Hood Method (Continued) –Re-establish threshold –If there is no change, increase masking by 5 dB –Re-establish threshold –If there is no change, record as threshold
AAO-HNS/F
Certificate Program for Otolaryngology Personnel Workshop
25
Why use the Hood Method? Goal of audiogram—gain ear specific information
Goal of plateau—establish level at which the threshold remains unchanged in the presence of a changing masking level
AAO-HNS/F
Certificate Program for Otolaryngology Personnel Workshop
26
Masking Examples –Air Conduction
Do you need to mask? Is the difference between ears 40 dB?
NO NO
250 Hz 500 Hz
15 – 10 40? 20 – 15 40?
NO
1000 Hz
30 – 20 40?
YES NO YES NO
2000 Hz 3000 Hz 4000 Hz 6000 Hz 8000 Hz
70 – 35 40? 80 – 40 40? 90 – 55 40? 90 – 40 40? 90 – 30 40?